HCPL030 High Current Pulse Source Meter
The HCPL030 is a pulse constant current source with features such as output pulse rising time 10μs, high test efficiency (40ms, with external control relay), support for two-channel pulse voltage measurement (peak sampling), and output polarity switching. Single output current is 300A and supports at least six multiple devices in parallel connection.
Basic Specification
| Item | Specification |
|---|---|
| Current Pulse Width | 50μs~1ms |
| Output Polarity Selection | Positive, Negative |
| Mini. Pulse Repetition Time | 100ms |
| Rising Time of Pulse Current | 10μs (The time of target current increasing from 10% to 90%) |
| Output Load Voltage | 20V@300A & pulse≤500μs |
| Interface | RS232、LAN |
| Noise | <65dB |
DUT Sampling
| Item | Specification |
|---|---|
| Measuring Range | 300mV 3V 30V |
| Resolution | 16Bit |
| Accuracy | 300mV Measuring range,±0.1%±1mV 3V Measuring range,±0.1%±3mV 30V Measuring range,±0.1%±8mV |
DUT Voltage Measurement
| Item | Specification |
|---|---|
| Independent Channel No. | 2 independent channel |
| Measurement Method | Remote measurement, peak voltage measurement (Sampling points can be adjustable) |
Output of Pulse Current
| Item | Specification |
|---|---|
| Measuring Range | 5A 100A 300A |
| Resolution | 16Bit |
| Accuracy | 5A Measuring range,±0.1%±16mA 100A Measuring range,±0.1%±128mA 300A Measuring range,±0.1%±256mA |
Current Readback
| Item | Specification |
|---|---|
| Measuring Range | 5A 100A 300A |
| Resolution | 16Bit |
| Accuracy | 5A Measuring range,±0.1%±16mA 100A Measuring range,±0.1%±128mA 300A Measuring range,±0.1%±256mA |