PLT2302 High Power Pulse Source Meter
The PLT2302 source meter is designed for testing visible light lasers, high-power BARs, VCSEL arrays, and other high-power laser devices. It supports pulse and DC PIV characteristic curve scanning, parameter calculation, and spectral measurements (requires an external spectrometer).
The instrument integrates a pulse current source, ammeter, voltmeter, and optical power meter, offering a compact size, fast testing speed, and ease of use.
Features
- Support us level pulse, DC output and test.
- Supports DC/Pulse mode. DC output can up to 10A/20W and pulse mode can up to 30A/20W.
- Min. Pulse width support 5us.
- Support Supports voltage 2-wire and 4-wire measurement, to maximize measurement accuracy.
- Support multi-devices combination, to provide higher output current.
- Supports high-speed acquisition and drawing of any single pulse waveform under continuous pulses, and real-time monitoring of pulse waveforms.
- Supports 2-channel sync pulse trigger output, with adjustable polarity and sync pulse width.
- Supports output polarity switching to adapt to different pin devices.
- Supports output short-circuit protection and over-temperature protection.
- PD response wavelength range: 850~1700nm.
Technical Specifications
Electrical Performance Spec.
DC Performance
Item Spec. Output Current 0~0.2A,±(0.05%RDG+0.1mA);
0~1A,±(0.05%RDG+0.25mA);
0~5A,±(0.05%RDG+1.25mA);
0~10A,±(0.05%RDG+2.5mA)Output Voltage 14V~26V Current Measurement 0~0.2A,±(0.05%RDG+0.05mA);
0~1A,±(0.05%RDG+0.25mA);
0~5A,±(0.05%RDG+1.25mA);
0~10A,±(0.05%RDG+2.5mA)Voltage Measurement 0~5V,±(0.05%RDG+1.25mV);
0~10V,±(0.05%RDG+2.5mV);
0~20V,±(0.05%RDG+5mV)Pulse Performance
Item Spec. Output Current 0~0.2A,±(0.05%RDG+0.05mA);
0~1A,±(0.05%RDG+0.25mA);
0~5A,±(0.05%RDG+1.25mA);
0~10A,±(0.05%RDG+2.5mA);
0~30A,±(0.05%RDG+7.5mA)Max. Output Voltage 26V Current Measurement 0~0.2A,±(0.05%RDG+0.05mA);
0~1A,±(0.05%RDG+0.25mA);
0~5A,±(0.05%RDG+1.25mA);
0~10A,±(0.05%RDG+2.5mA);
0~30A,±(0.05%RDG+7.5mA)Voltage Measurement 0~5V,±(0.05%RDG+1.25mV);
0~10V,±(0.05%RDG+2.5mV);
0~20V,±(0.05%RDG+5mV)Pulse Rising and Falling time ≤20% x Pulse Width Pulse Width Duty Ratio D≤360/(I×U)×100%
[Note] D is the duty ratio, the max. doesn't exceed 100%;
I is the load current (unit: A); U is the output voltage (unit: V).Pulse Width 5us~500ms Optoelectronic Performance Spec.
Optical Port Power
Item Spec. Optical Port Power Detection Range and Accuracy 0~100uW,±(0.05%RDG+0.03uW);
0~1mW,±(0.05%RDG+0.25uW);
0~5mW,±(0.05%RDG+1.25uW)Electrical Port Current
Item Spec. Electrical Port Current Detection Range and Accuracy 0~100uA,±(0.05%RDG+0.03uA);
0~1mA,±(0.05%RDG+0.25uA);
0~10mA,±(0.05%RDG+2.5uA)MPD
Item Spec. MPD + Reverse Bias Voltage 0~5V,±(1%RDG+50mV) Current Detection Range and Accuracy 0~100uA,±(0.05%RDG+0.03uA);
0~1mA,±(0.05%RDG+0.25uA);
0~10mA,±(0.05%RDG+2.5uA)
Measured Data Example